Development and Evaluation of an Improved Method for Screening of Amphetamines*
J. Shindelman, J. Mahal, G. Hemphill, P. Pizzo, and W.A.
Coty
We developed a homogeneous immunoassay method to eliminate false-positive amphetamine
results caused by cross-reactive substances, including over-the-counter allergy
and cold medications. This method uses a neutralizing antibody that binds to amphetamines
but does not bind to the labeled amphetamine conjugate used in the assay. The
amount of neutralizing antibody is sufficient to reduce the assay signal resulting
from authentic amphetamine and methamphetamine, but not the signal resulting from
cross-reactants. This concept was implemented using the CEDIA® DAU Amphetamines
assay on Hitachi 747 and 717 clinical chemistry analyzers. Urine samples were
tested using the standard, unmodified reagents in one channel and reagents containing
the neutralizing antibody in a second channel. The difference in rate between
the two tests was calculated by the analyzer; true-positive samples showed a significantly
greater decrease in assay signal in response to neutralizing antibody as compared
with false-positive samples. The neutralization method was evaluated in two studies
using 448 samples that tested positive in the initial CEDIA DAU Amphetamines screening
test. The samples were separated into categories of 154 true-positive samples
and 294 false-positive samples based upon a secondary screen with the Abbott FPIA
Amphetamines assay followed by gas chromatographymass spectrometry (GCMS)
testing using the HHS (SAMHSA) cutoff criteria. The CEDIA neutralization test
successfully identified all 154 of the GCMS confirmed positive samples.
The test successfully identified as false positive 251 out of the 294 (85.4%)
samples that failed to confirm by GCMS.
Reproduction of editorial content
of this journal is prohibited without publishers
permission.
This article is available in its entirety by fax for $40.00 each.
Visa, MasterCard and AMEX accepted.
To order electronically click
here
or call: 847-647-2900 ext. 1323
or fax request to: 847-647-1155.
Please indicate JAT
volume and issue along with page numbers.
|
| Home
| Subscribe | Current Issue
| Back Issues | Search | Advertise | Other Publications
| Site Map |