Journal of Analytical Toxicology Article Abstracts

Journal of Analytical Toxicology Horizontal Line

Published: Journal of Analytical Toxicology, Volume 23, Number 7, Nov/Dec 1999, pp. 620-624

Determination of Opiates and Cocaine and Its Metabolites in Biological Fluids by High-Performance Liquid Chromatography with Electrospray Tandem Mass Spectrometry
A. Cailleux, A. Le Bouil, B. Auger, G. Bonsergent, A. Turcant, and P. Allain

A rapid, sensitive, and specific method for the determination of opiates and cocaine and metabolites in urine, plasma, and blood was established. A one-step extraction followed by liquid chromatography–electrospray ionization tandem mass spectrometry operating in multiple reaction monitoring mode was used. Two chromatographic runs were performed, each in less than 6 min. The lower limit for accurate quantitative determination was 5 µg/L for cocaine and metabolites and 10 µg/L for opiates. Linearity was obtained from 10 to 1000 µg/L. Intraday (n = 6) and interday (n = 6) precisions and recoveries (n = 6) were determined at 10 or 25, 100, and 1000 µg/L concentrations. Precisions with a coefficient of variation less than 15% were obtained. Recoveries between 85 and 115% were determined.

Reproduction of editorial content of this journal is prohibited without publisher’s permission.

This article is available in its entirety by fax for $40.00 each.
Visa, MasterCard and AMEX accepted.

To order electronically click here
or call: 847-647-2900 ext. 1323
or fax request to: 847-647-1155.

Please indicate JAT volume and issue along with page numbers.