Journal of Analytical Toxicology Article Abstracts

Journal of Analytical Toxicology Horizontal Line

Published: Journal of Analytical Toxicology, Volume 23, Number 3, May/June 1999, pp.137-140.

The Measurement of Nitrite in Adulterated Urine Samples by High-Performance Ion Chromatography
Jasbir Singh, James A. Elberling, D. Gary Hemphill, and Jeffrey Holmstrom

With the increased availability of nitrite-containing commercial products that are used for the adulteration of urine samples in the workplace, it is necessary for laboratories to be able to detect and confirm the presence of nitrite in these samples. We have developed a method to confirm the presence of nitrite in urine samples. The method uses the IonPac AS 14 analytical column with the Dionex series 4500I Bio LC system equipped with an anion self-generating suppressor and conductivity detector. Using a single-point calibration, the method is linear and accurately quantitates nitrite to 12,000 µg/mL. The limit of detection is 30 µg/mL, and the day-to-day precision of the assay has a coefficient of variation (CV) of 4.3% at 1200 µg/mL and 3.8% at 2700 µg/mL of nitrite.

Reproduction of editorial content of this journal is prohibited without publisher’s permission.

This article is available in its entirety by fax for $4.00 per page.
Visa or MasterCard accepted.

To order electronically click here
or call: 847-647-2900 ext. 1323
or fax request to: 847-647-1155.

Please indicate JAT volume and issue along with page numbers.

 