Published: Journal of Analytical Toxicology, Volume 22, Number 1, January/February 1998, pp. 2732.
Diagnostic Performance
of the EMIT-tox® Benzodiazepine Immunoassay, FPIA
Serum Benzodiazepine Immunoassay, and Radioreceptor Assay in
Suspected Acute Poisoning
A.G. Verstraete, F.M. Belpaire, and G.G. Leroux-Roels
We evaluated the diagnostic performance of the EMIT-tox serum benzodiazepine assay adapted to a Hitachi 717 analyzer (EMIT), the Abbott ADx serum benzodiazepine fluorescence polarization immunoassay (FPIA), and a radioreceptor assay (RRA) in 113 patients with suspected acute poisoning. The reference method was high-performance liquid chromatography with ultraviolet detection after solid-phase extraction. For the discrimination between negative and positive samples, the areas under the receiver-operating characteristic (ROC) curves were 0.976, 0.991, and 0.991 for EMIT (cutoff, 50-ng/mL diazepam), FPIA (cutoff, 12-ng/mL nordiazepam), and RRA (cutoff, 50-ng/mL diazepam), respectively. For the discrimination between non-toxic and toxic concentrations, the areas under the ROC curves were 0.896, 0.893, and 0.933, respectively. EMIT (with the cutoff lowered to 50 ng/mL), FPIA, and RRA can be reliably used to screen for the presence of benzodiazepines in serum, but in many cases they cannot discriminate between toxic and nontoxic concentrations.
Reproduction of editorial content of this journal is prohibited without publishers permission.