Published: Journal of Analytical Toxicology, Volume 21, Number 7, November/December 1997, pp. 521–528.

The Analysis of EDTA in Dried Bloodstains by Electrospray LC–MS–MS and Ion Chromatography
M.L. Miller, B.R. McCord, R. Martz, and B. Budowle

Analytical methods were developed to determine the presence of ethylenediaminetetraacetic acid (EDTA) in dried bloodstains to provide probative information when allegations of evidence tampering have been made in criminal cases. A simple screening method using ion chromatography to analyze stains was found to be quantitative to the 5 ppm level. The presence of EDTA was then confirmed using negative and positive ion mode liquid chromatography–tandem mass spectrometry (LC–MS–MS) methods. A blind trial of these methods on 42 samples correctly determined the bloodstains that did and did not contain the preservative EDTA. One interesting observation in these results was the adsorption and postanalysis release of EDTA in the chromatographic system. In order to avoid cross contamination of samples resulting from this phenomena, it was found to be necessary to use EDTA-free blood extracts as blanks in the LC–MS analysis of bloodstains.

Reproduction of editorial content of this journal is prohibited without publisher’s permission.

This article is available in its entirety by fax for $40.00 each.
Visa, MasterCard and AMEX accepted.

To order electronically click here
or call: 847-647-2900 ext. 1323
or fax request to: 847-647-1155.

Please indicate JAT volume and issue along with page numbers.

 